KP-700 Multi-parameter Veterinary Monitor
Model KP-700
A 12.1-inch veterinary patient monitor that tracks ECG, NIBP, SpO2, pulse rate, temperature and respiration as standard, with EtCO2 available for anesthesia monitoring.
- 12.1" high-resolution display showing up to 9 waveforms, including 7-channel ECG at once
- 13 types of arrhythmia analysis, real-time ST-segment analysis and pacemaker detection
- 720 hours of trend and graphic review, 2 hours of waveform review, 1,000 NIBP groups and 200 alarm events
- Three-level audible and visual alarms with voice prompts, alarm review and pause
Overview
The KP-700 runs on an ARM9 main board with embedded Linux. Six display layouts (standard, big font, trend table, full-lead ECG, NIBP trend and OxyCRG) let each team choose what they see first.
Built-in tools include drug calculation with titration tables, hemodynamic calculation, an NIBP self-test mode and multi-language display. Patient data exports over USB.
Alarms cover sensor-off and paper-out conditions as well as physiological limits, with three alarm levels, voice prompts, alarm review and alarm pause.
Configuration
| Standard | 12.1" display, 3/5-lead ECG, NIBP, SpO2, pulse rate, temperature, respiration |
|---|---|
| Optional | EtCO2, Masimo® SpO2, touch screen, central monitoring station, wall mounting |
Specifications
ECG
| Lead mode | 3/5 leads: I, II, III / I, II, III, aVR, aVL, aVF, V |
|---|---|
| Heart rate range | 15–350 bpm |
| Accuracy | ±1% |
| Resolution | 1 bpm |
| Gain | ×0.25, ×0.5, ×1, ×2 |
| Sweep speed | 12.5, 25, 50 mm/s |
| Signal range | ±5 mV p-p |
| Bandwidth | Monitor 0.5–40 Hz; Diagnostic 0.05–130 Hz |
| Protection | Breakdown voltage 4000 VAC, 50/60 Hz |
| Leakage current | < 10 µA |
NIBP
| Method | Oscillometric |
|---|---|
| Modes | Manual, Auto, STAT |
| Auto intervals | 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 min |
| STAT cycle | 5 min |
| Systolic | 40–280 mmHg |
| Mean | 20–240 mmHg |
| Diastolic | 10–220 mmHg |
| Static pressure accuracy | ±3 mmHg |
| Accuracy | Max. mean error ±5 mmHg; max. SD 8 mmHg |
| Resolution | 1 mmHg |
| Overpressure protection | Dual |
SpO2
| Range | 0–100% |
|---|---|
| Resolution | 1% |
| Accuracy | ±2% (70–100%); 0–69% unspecified |
| Audio | Pitch tone, multi-level volume |
| Waveform speed | 12.5, 25 mm/s |
Pulse rate
| Range | 20–250 bpm |
|---|---|
| Accuracy | ±3 bpm |
| Resolution | 1 bpm |
Respiration
| Method | Impedance, RA-LL / RA-LA |
|---|---|
| Rate | 0–150 BrPM |
| Accuracy | ±2 BrPM or ±2%, whichever is greater |
| Resolution | 1 BrPM |
| Gain | ×0.25, ×0.5, ×1, ×2, ×4 |
| Sweep speed | 6.25, 12.5, 25 mm/s |
| Apnea alarm | 10–40 s |
Temperature
| Technique | Thermistor probe (2.25K) |
|---|---|
| Channels | 1 |
| Range | 0–50 °C |
| Resolution | 0.1 °C |
| Measuring time | < 10 s |
EtCO2 (optional)
| Range | 0–100 mmHg |
|---|---|
| Resolution | 0.1 mmHg |
| Accuracy | 0–40 mmHg: ±2 mmHg; 41–70: ±5% of reading; 71–100: ±8% of reading |
| Respiratory rate | 3–150 BrPM, ±1 |
General
| Power | AC 100–240 V, 50/60 Hz |
|---|---|
| Battery | 4 h working time |
| Operating environment | 5–40 °C, 15–90% RH |
| Safety | IEC 60601-1 |
Specifications may change without notice. Contact us to confirm the configuration for your order.
Request a quote
Get pricing, configuration options and lead time for the KP-700 Multi-parameter Monitor.
Bao'an District, Shenzhen, Guangdong, China