Anesthesia machine

VET-60 Veterinary Anesthesia Machine

Model VET-60

An ergonomic anesthesia and ventilation system for pet hospitals, research institutes and teaching labs. Enter the animal's weight and the built-in database suggests tidal volume, I:E ratio, respiratory rate and pressure, ready to fine-tune for the case.

  • Four ventilation modes: VCV, PCV, SIMV-VC and SPONT, plus manual and ACGO
  • 10.1" display with real-time flow and pressure waveforms, tidal volume and end-tidal data
  • Interchangeable bellows (0–300 mL and 0–1,500 mL) for small and large patients
  • Vaporizer with pressure, flow and temperature compensation for isoflurane or sevoflurane
20–1,500 mLTidal volume range
2–60 BPMRespiratory rate
≥150 minBattery backup

Ventilation modes

VCV

Volume control

Inspiration ends once the delivered gas volume reaches the set tidal volume.

PCV

Pressure control

Inspiration ends once airway pressure reaches the set peak inspiratory pressure.

SPONT

Spontaneous

If spontaneous breathing stops or weakens for a set time, the machine steps in with assisted breaths, then hands back control when the animal recovers.

SIMV

Synchronized intermittent mandatory

Mandatory breaths are synchronized with the animal's own effort within a trigger window, with pressure support for spontaneous breaths outside it.

Alarms you can act on

Set limits for peak pressure, tidal volume, minute volume, rate, PEEP, oxygen concentration, end-tidal CO2 and apnea.

Three-level, color-coded audible and visual alarms make priority clear at a glance. Alarm volume is adjustable, every alarm can be enabled or disabled, and alarm records can be reviewed after the procedure.

Designed for daily use

The CO2 absorber canister has a handle for one-handed removal. A single switch moves between open and closed circuit, and another between manual and mechanical ventilation. The APL valve carries a precise pressure scale, and the Selectatec®-type bypass mount makes vaporizer installation and maintenance simple.

Removing the CO2 absorber canister with one hand
One-hand absorber removal
ACGO switch on the VET-60 front panel
One-key open/closed circuit switch
Small and large bellows
Interchangeable bellows
APL valve knob with scale
APL valve with pressure scale

Specifications

Main specifications

Display10.1" (1280 × 800)
PowerAC 100–240 V, 50/60 Hz
Battery≥ 150 min
Gas supply pressure0.28–0.6 MPa
O2 flush25–75 L/min
Flowmeter0–4 L/min
APL valve1–70 cmH2O
CO2 absorbent1,200 mL
Airway pressure−20 to 100 cmH2O
ACGOOne-key switching
Bellows0–300 mL / 0–1,500 mL

Ventilator (gas-driven, electronically controlled)

ModesVCV, PCV, SIMV-VC, SPONT, Manual, ACGO
Tidal volume20–1,500 mL
Rate2–60 BPM
I:E4:1 to 1:8
PEEPOFF / 3–30 cmH2O
Pinsp5–50 cmH2O
Inspiratory hold≤ 30 s
Tinsp0.2–10 s
Plimit10–60 cmH2O
Flow triggerOFF / 1–20 L/min
Apnea time15–30 s
Monitored valuesVT, MV, Ppeak, PEEP, Rate, EtCO2, FiCO2, FiO2

Standard configuration

VaporizerPressure, flow and temperature compensated
Isoflurane0.2–6%
Sevoflurane0.2–8%
Vaporizer mountSingle
Oxygen hoseNIST
Breathing circuit22 mm / 15 mm
Breathing bags0.5 / 1 / 2 / 3 L

Optional

FlowmeterDual gas (O2, Air), 0–5 L/min
CO2 module0–100 mmHg
O2 sensor21–100%
Endotracheal tubesSizes 2–10
LaryngoscopeBlades 2, 3, 4 (LED)
Anesthesia masksSizes 4, 5, 6
Non-rebreathing circuitAvailable

Specifications may change without notice. Contact us to confirm the configuration for your order.

Request a quote

Get pricing, configuration options and lead time for the VET-60 Anesthesia Machine.

Shenzhen Roeso Technology Co., Ltd.
Skyworth Digital Building, Shiyan Subdistrict,
Bao'an District, Shenzhen, Guangdong, China